Wave Logger
Compatible with external sensor
Core Features
Product Application
Determine the product's adherence to your specifications.
OSSI-010-004E Wave Logger & OSSI-010-005xE RF Wave Logger – Product Description
The Ocean Sensor Systems Wave Logger and RF Wave Logger are high-resolution, self-powered instruments designed for direct-contact measurement of liquid surface height at data rates of up to 30 Hz, ideal for waves, wakes, tides, and liquid level monitoring in various environments.
Measurement & Performance
| Measurement Principle | Direct contact conductive staff measurement |
|---|---|
| Sampling Rate | Up to 30 Hz |
| Staff Length Options | Interchangeable staffs available in a variety of lengths |
| Output Range | 0-4095 counts, convertible to millimeters |
| Accuracy | Better than 1% of full scale (0.1% with advanced post-processing) |
Data Logging & Communication
| Data Storage | Internal logging to Compact Flash memory cards (64MB to 2GB) |
|---|---|
| Data Format | PC readable format for easy analysis |
| Wired Communication | RS232 serial communication for configuration and real-time data |
| Wireless Option (RF Model) |
900MHz RF Wave Logger: X09-009NSC modem 2.4GHz RF Wave Logger: X24-009NSC modem Compatible with Stand-Alone RF Modems for remote data links |
| Configuration Software | Graphical User Interface available for download |
Power & Battery
| Power Source | Replaceable internal battery pack of 4 alkaline C-cells |
|---|---|
| Battery Replacement | Individual alkaline C cells connected in series |
| Battery Status | Replace when voltage is at or below 3.8V |
| Battery Life Indicator | Open circuit reading of 6V or greater indicates ≥90% capacity |
Construction & Deployment
| Housing | Waterproof housing with solid-state electronics |
|---|---|
| Materials | Glass, PTFE, stainless steel, and ABS plastic |
| Mounting | Vertical orientation with staff projecting downward into fluid |
| Environmental Protection | No special protection needed from spray, rain, or sun |
| Mounting Hardware | Two 10-24 (3/4” deep) mounting holes on base; Optional Mounting Hardware Kit (OSSI-015-002) |
| Grounding Requirement | Electrical connection from ground stud to fluid being measured |
Operation & Indicators
| Startup | Automatically begins data collection within 60 seconds of power-up |
|---|---|
| LED Indicator |
3-second light: Normal operation Rapid blinking (~3Hz): CF card error Slow blinking (1Hz): Safe to remove CF card/battery |
| Card Removal | Momentarily depress button and wait for slow LED blink before removal |
| Fluid Compatibility | Almost any fluid compatible with construction materials |
Control & Interface
| Interface Software | Staff & Logger Interface Software for Windows |
|---|---|
| Configuration Options | Sample rate, data format, start times, file intervals, burst sampling |
| Alternative Interface | HyperTerminal program with 9600 baud, no flow control |
| RF Configuration | Configure via AT commands or Digi International's X-CTU software |
Maintenance
| Routine Maintenance | Wiping slime buildup off the yellow PTFE staff |
|---|---|
| Staff Replacement | Staff can be separated from electronics housing for replacement |
| Battery Check | Primary maintenance is checking battery voltage |
| Calibration | Required when staff is changed; otherwise minimal maintenance needed |
Typical Applications
| Wave & Wake Measurement | Ocean waves and vessel wake characterization |
|---|---|
| Water Level Monitoring | Tides, lake levels, pond levels, tank levels, pool levels |
| Industrial Applications | Sinkage measurement, trim measurement for vessels |
| Research | Coastal research, environmental monitoring, hydrological studies |
| Key Advantage | Robust direct-contact sensing with reliable internal data logging for long-term deployment |
Similar Products
You might find a more suitable solution in one of these products.
Wave Logger III vs Wave Logger – Comparison Table
Key differences between the third-generation Wave Logger III and the earlier Wave Logger model
Product Overview
| Feature | Wave Logger III (OSSI-010-010) | Wave Logger (OSSI-010-004E/005xE) |
|---|---|---|
| Product Generation | Third Generation | Earlier Generation |
Grounding & Installation
| Grounding System | Self-grounding stud - simplified installation | External grounding wire required - must connect ground stud to water with copper wire |
|---|---|---|
| Staff Grounding Types | Internal Ground (Int Gnd) options available | External Ground (Ext Gnd) primarily used |
Technical Specifications
| Measurement Performance | Up to 30Hz sampling rate | Up to 30Hz sampling rate |
|---|---|---|
| Data Storage | Compact Flash cards (64MB-2GB) | Compact Flash cards (64MB-2GB) |
| Power System | 4 C-cell alkaline batteries | 4 C-cell alkaline batteries |
Design & Physical Features
| Staff Mounting & Clearance | Blue staff section - keep 4-6" from metallic surfaces | Yellow staff section - keep at least 4" from metallic surfaces |
|---|---|---|
| Staff Types | Coaxial Rod Staff & Coaxial Cable Staff options | Rod and Cable Staff options |
| Waterproof Housing | Yes - no special protection needed | Yes - no special protection needed |
Performance & Accuracy
| Accuracy | 1% full scale (0.1% with calibration) | Better than 1% full scale (0.1% with calibration) |
|---|---|---|
| Calibration Procedure | Simplified grounding with internal ground options | External grounding required during calibration |
Software & Control
| Software Compatibility | Wave Logger III Interface Software | Wave Staff/Logger Interface Software |
|---|---|---|
| Communication Options | RS232 standard, RF optional | RS232 standard, RF optional |
Maintenance & Service
| Staff Interchangeability | Yes - multiple lengths available | Yes - multiple lengths available |
|---|---|---|
| Routine Maintenance | Wipe slime buildup off yellow PTFE staff | Wipe slime buildup off yellow PTFE staff |
Key Improvements in Wave Logger III:
- Enhanced Grounding: Self-grounding system eliminates need for external grounding wires
- Improved Staff Design: Coaxial staff options with internal grounding provide better performance
- Simplified Installation: Reduced setup complexity and hardware requirements
- Updated Design: Blue staff section and refined mechanical design